header
Image from OpenLibrary

Validation and clinical implementation of COMPASS system 3D dose verification systems for quality assurance in volumetric-modulated arc therapy techniques / by Samar Emad Mohamed Salem; Prof. Dr. Reem Elgebaly, Dr. Noha Gamal Al-Din, Dr. Hany Ammar.

By: Contributor(s): Material type: TextTextLanguage: English Summary language: English, Arabic Producer: 2023Description: 65 pages : illustrations ; 25 cm. + CDContent type:
  • text
Media type:
  • Unmediated
Carrier type:
  • volume
Other title:
  • التاكد والتصديق الطبي علي استخدام نظام كومباس ك أجهزة تحقق ثلاثية العلاج الاشعاعي الحجمي القوسي متغير الشدة [Added title page title]
Subject(s): DDC classification:
  • 615.842 21
Online resources: Available additional physical forms:
  • Issues also as CD.
Dissertation note: Thesis (M.Sc.)-Cairo University, 2023. Summary: Compass system was prepared for the clinical measurements through collecting the required data for the compass collapse cone algorithm for VersaHD machines and using the beam modeling. By using Compass dose verification provide three dimensional (3D) measured dose reconstruction on the same computed tomography (CT) scan used in the planning, where the conventional two dimensional (2D) detector system doesn’t give information regards the errors dose delivery in planning target volume (PTV) and organs at risk (OARs). In addition the Compass reconstructed dose provide dose volume histogram (DVH) for the different structures compared to the reference treatment planning system (TPS)-DVH. The COMPASS QA system uses MatriXX detector with gantry angle sensor. It consists of two parts; COMPASS software and an associated measuring device. COMPASS software is able to recalculate and reconstruct a 3D dose distribution on the CT data of the patient based on measured fluence from MatriXX detector of the patient based on measured fluence from MatriXX detector and collapsed cone convolution/superposition algorithm. The aim of the work commissioning and Clinical Validation of the COMPASS system for two matched Elekta Versa HD Linear Accelerators by comparing computation dose and reconstruction dose in each VMAT plans of statistically acceptable number of different anatomical site patients. The doses and DVHs computed using Collapsed Cone Algorithm of the COMPASS system were compared with Monaco TPS calculated plans using Monte Carlo.Summary: الهدف من العلاج الاشعاعي توصيل الجرعه المطلوبه للانسجه المصابه بالورم وتجنب الانسجه السليمه وهذا مايحققه التقنيه الحديثه في العلاج الاشعاعي وهي العلاج الحجمي متغير الشده . حيث ان العلاج الحجمي متغير الشده يوفر توزيع متجانس ومحدد من الجرعه علي الورم وتحمي باقي الاعضاء من الاشعاع ونستطيع تحقيق هذا بواسطه تقنيات مختلفه مثل العلاج الشعاعي متغير الشده بتكوين اقواس من الاشعاع بواسطه تغيير زوايا موجهه الاشعاع وسرعة موجة الاشعاع وتغير معدل الاشعاع و محدد متعدد الاصابع ولوجود هذه التقنيات المعقده لذا يجب ان نتاكد من وصول الجرعه الي المكان المحدد من قبل تطبيقه علي المريض . ومن الاساليب التقليديه من التحقق من جوده الخطه العلاجيه بتطبيق نفس الخطه علي فانتوم وقياس الجرعه بواسطه حجره ايونيه وحديثا اصبحوا يستخدموا صوره الكترونيه ومصفوفه ثنائيه الابعاد ليوضح توزيع الجرعه الاشعاعيه للعلاج بواسطه العلاج الحجمي متغير الشده والعلاج متغير الشده ولكن ينقص هذه التقنيه توضيح الجرعه التي يتعرض ليها باقي الاعضاء الموجوده في مسار الاشعه والقريبه من الحقل الاشعاعي وهذا الذي تغلب عليه جهاز كومباس . حيث ان هذا الجهاز يتاكد من الجرعه في الثلاثه ابعاد ويستطيع ايجاد وحساب الجرعه التي يتعرض اليها كل عضو من اعضاء الجسم . يستخدم هذا الجهاز ماتركس و جهاز استشعاري لزاويه موجهه الاشعاع ويتكون هذا الجهاز من برنامج يتم فيه اعاده قياس الاشعاع وايضا اعاده بناء الحسابات بناءا علي توزيع الاشعاع في ماتركس والمعلومات من الاشعه المقطعيه وحسابها بنظام لوغرتمي مختلف. الهدف Elekta الهدف من البحث التحقق من تشغيل جهازكومباس علي جهازين من اجهزه المعجل الخطي الكتا فرسا اتش دي( ومقارنه خطه العلاج الاشعاعي باستخدام تقنيه العلاج الاشعاعي (versa HD الحجمي متغير الشده وذلك في عدد من الحالات لديهم اورام في اماكن مختلفه للتاكد من جوده خطه العلاج الاشعاعي
Tags from this library: No tags from this library for this title. Log in to add tags.
Star ratings
    Average rating: 0.0 (0 votes)

Thesis (M.Sc.)-Cairo University, 2023.

Bibliography: pages 59-65.

Compass system was prepared for the clinical measurements
through collecting the required data for the compass collapse cone
algorithm for VersaHD machines and using the beam modeling.
By using Compass dose verification provide three dimensional
(3D) measured dose reconstruction on the same computed
tomography (CT) scan used in the planning, where the
conventional two dimensional (2D) detector system doesn’t give
information regards the errors dose delivery in planning target
volume (PTV) and organs at risk (OARs). In addition the
Compass reconstructed dose provide dose volume histogram
(DVH) for the different structures compared to the reference
treatment planning system (TPS)-DVH. The COMPASS QA
system uses MatriXX detector with gantry angle sensor. It
consists of two parts; COMPASS software and an associated
measuring device. COMPASS software is able to recalculate and
reconstruct a 3D dose distribution on the CT data of the patient
based on measured fluence from MatriXX detector of the patient
based on measured fluence from MatriXX detector and collapsed
cone convolution/superposition algorithm. The aim of the work
commissioning and Clinical Validation of the COMPASS system
for two matched Elekta Versa HD Linear Accelerators by
comparing computation dose and reconstruction dose in each
VMAT plans of statistically acceptable number of different
anatomical site patients. The doses and DVHs computed using
Collapsed Cone Algorithm of the COMPASS system were
compared with Monaco TPS calculated plans using Monte Carlo.

الهدف من العلاج الاشعاعي توصيل الجرعه المطلوبه للانسجه المصابه بالورم وتجنب الانسجه السليمه
وهذا مايحققه التقنيه الحديثه في العلاج الاشعاعي وهي العلاج الحجمي متغير الشده .
حيث ان العلاج الحجمي متغير الشده يوفر توزيع متجانس ومحدد من الجرعه علي الورم وتحمي باقي
الاعضاء من الاشعاع ونستطيع تحقيق هذا بواسطه تقنيات مختلفه مثل العلاج الشعاعي متغير الشده
بتكوين اقواس من الاشعاع بواسطه تغيير زوايا موجهه الاشعاع وسرعة موجة الاشعاع وتغير معدل
الاشعاع و محدد متعدد الاصابع ولوجود هذه التقنيات المعقده لذا يجب ان نتاكد من وصول الجرعه الي
المكان المحدد من قبل تطبيقه علي المريض .
ومن الاساليب التقليديه من التحقق من جوده الخطه العلاجيه بتطبيق نفس الخطه علي فانتوم وقياس
الجرعه بواسطه حجره ايونيه وحديثا اصبحوا يستخدموا صوره الكترونيه ومصفوفه ثنائيه الابعاد
ليوضح توزيع الجرعه الاشعاعيه للعلاج بواسطه العلاج الحجمي متغير الشده والعلاج متغير الشده
ولكن ينقص هذه التقنيه توضيح الجرعه التي يتعرض ليها باقي الاعضاء الموجوده في مسار الاشعه
والقريبه من الحقل الاشعاعي وهذا الذي تغلب عليه جهاز كومباس .
حيث ان هذا الجهاز يتاكد من الجرعه في الثلاثه ابعاد ويستطيع ايجاد وحساب الجرعه التي يتعرض اليها
كل عضو من اعضاء الجسم . يستخدم هذا الجهاز ماتركس و جهاز استشعاري لزاويه موجهه الاشعاع
ويتكون هذا الجهاز من برنامج يتم فيه اعاده قياس الاشعاع وايضا اعاده بناء الحسابات بناءا علي توزيع
الاشعاع في ماتركس والمعلومات من الاشعه المقطعيه وحسابها بنظام لوغرتمي مختلف.
الهدف
Elekta الهدف من البحث التحقق من تشغيل جهازكومباس علي جهازين من اجهزه المعجل الخطي
الكتا فرسا اتش دي( ومقارنه خطه العلاج الاشعاعي باستخدام تقنيه العلاج الاشعاعي (versa HD
الحجمي متغير الشده وذلك في عدد من الحالات لديهم اورام في اماكن مختلفه للتاكد من جوده خطه
العلاج الاشعاعي

Issues also as CD.

Text in English and abstract in Arabic & English.

There are no comments on this title.