Study of physical and technical factors associated with small and small shaped MLC photon beam in recent radiotherapy techniques / Dina Mohamed Abdelaziz Ibrahim ; Supervised Nader Elsherbini , Wafaa Khalel

By: Contributor(s): Material type: TextTextLanguage: English Publication details: Cairo : Dina Mohamed Abdelaziz Ibrahim , 2015Description: 107 P. : charts , facsimiles ; 25cmOther title:
  • دراسة استخذام العوامل الفيزيائية و التكنيكية المرتبطة بالحقول الإشعاعية الصغيرة بمحدد الأشعة متعدد الوريقات في تقنيات العلاج الاشعاعي الحديثة [Added title page title]
Subject(s): Online resources: Available additional physical forms:
  • Issued also as CD
Dissertation note: Thesis (Ph.D.) - Cairo University - Faculty of Science- - Department of Biophysics Summary: n this study a dosimetric measurements are done for IMRT in the buildup region with different measurement tools (Gafchromic films, parallel plate ion chamber and TPS) at different gantry angle to assess the discrepancies in this region figure (3-1) and table (3-1).The study provide also the most accurate collimation method in the small fields either MLC only or MLC+Jaws to assess which of them can help the treatment planning system tabulated dosimetric functions to introduce a precise dose calculation. In the same way assessments of the output factor and leaf transmission are measured. Technical and dosimetric parameters related to the use of small shaped MLC field size are investigated in terms of the evaluation of MONACO version 3.3.0 versus XIO version 5.0.0, biological and physical optimization and dynamic and static MLC, using the chair, penumbral and different doses tests. This study revealed that the beam data measurements that will be used in the modeling of small shaped MLC field sizes have to be measured by MLC only without jaws to mimic the real condition of using MLC in shaping small irregular fields. Also, the results indicate that MONACO biological optimization is better than XIO physical optimization. Dynamic plan delivery is time saving more than static plans with no side effect on the plan quality
Tags from this library: No tags from this library for this title. Log in to add tags.
Star ratings
    Average rating: 0.0 (0 votes)
Holdings
Item type Current library Home library Call number Copy number Status Barcode
Thesis Thesis قاعة الرسائل الجامعية - الدور الاول المكتبة المركزبة الجديدة - جامعة القاهرة Cai01.12.04.Ph.D.2015.Di.s (Browse shelf(Opens below)) Not for loan 01010110069654000
CD - Rom CD - Rom مخـــزن الرســائل الجـــامعية - البدروم المكتبة المركزبة الجديدة - جامعة القاهرة Cai01.12.04.Ph.D.2015.Di.s (Browse shelf(Opens below)) 69654.CD Not for loan 01020110069654000

Thesis (Ph.D.) - Cairo University - Faculty of Science- - Department of Biophysics

n this study a dosimetric measurements are done for IMRT in the buildup region with different measurement tools (Gafchromic films, parallel plate ion chamber and TPS) at different gantry angle to assess the discrepancies in this region figure (3-1) and table (3-1).The study provide also the most accurate collimation method in the small fields either MLC only or MLC+Jaws to assess which of them can help the treatment planning system tabulated dosimetric functions to introduce a precise dose calculation. In the same way assessments of the output factor and leaf transmission are measured. Technical and dosimetric parameters related to the use of small shaped MLC field size are investigated in terms of the evaluation of MONACO version 3.3.0 versus XIO version 5.0.0, biological and physical optimization and dynamic and static MLC, using the chair, penumbral and different doses tests. This study revealed that the beam data measurements that will be used in the modeling of small shaped MLC field sizes have to be measured by MLC only without jaws to mimic the real condition of using MLC in shaping small irregular fields. Also, the results indicate that MONACO biological optimization is better than XIO physical optimization. Dynamic plan delivery is time saving more than static plans with no side effect on the plan quality