Verification of performance of algorithm implemented in eclipse treatment planning system /
تقييم أداء طريقة الحساب المحملة على جهاز التخطيط العلاجى Eclipse
Mostafa Abdalmoez Ramadan ; Supervised Alsyed Awad Gafar , Nader Elsherbini
- Cairo : Mostafa Abdalmoez Ramadan , 2009
- 173Leaves : charts , facsimiles ; 30cm
Thesis (M.Sc.) - Cairo University - Faculty of Science - Department of Biophysics
This study is performed aiming to verify the pencil beam convolution algorithm implemented on Eclipse treatment planning system (from Varian ) version 7.3 . Elekta precise linear accelerator producing two photon beams of 6 and 15 MV are used . All measured depth doses and does profiles are scanned using Weihofer WP 700 water phantom and two cylindrical water proof chambers each of sensitive volume 0.147 cm and wall thickness of 0.4 mm
Dose calculation algorithms Eclipse Planning system