Investigation of cone-beam computed tomography (CBCT) in conventional & intense modulated radiotherapy (IMRT) / Adly Fawzy Abdelsalam Naga ; Supervised Anwar Abdelazeim Elsayed , Ehab Marouf Attalah , Mohamed Lotayef,

By: Contributor(s): Material type: TextLanguage: English Publication details: Cairo : Adly Fawzy Abdelsalam Naga , 2014Description: 94 Leaves : charts , facsimiles ; 25cmOther title:
  • بحث الاشعة المقطعية الالية ذات الشعاع المخروطى فى العلاج الاشعاعى التقليدى والمتغير الشدة [Added title page title]
Subject(s): Online resources: Available additional physical forms:
  • Issued also as CD
Dissertation note: Thesis (Ph.D.) - Cairo University - Faculty of Science - Department of Biophysics Summary: Purpose: Needs to employ quality assurance (QA) programs that ensure that the system performance meets high expectations consistent with patient care requirements. And to improve target localization accuracy, so that PTV treatment margin can be adjusted or reduced. Materials and Methods: Recently installed an Elekta Synergy linear accelerator with energy 6MV (Elekta, Synergy S, XVI, Crawley, UK) system. Catphan 503 phantom to evaluate CBCT image quality. Ball-Bearing Phantom to evaluate the geometric accuracy of kV and MV isocenters. Quasar{u2122} Penta-Guide Phantom for daily quality assurance (QA). Results: In the present study over a period of 12 months showed that, MV and kV isocenters with mean value of 0.10±0.29mm, -0.07±0.37mm, 0.04±0.28mm and the maximal deviation were 0.45mm, -0.55mm and -0.47mm in x (lateral), y (longitudinal) and z (height) respectively
Tags from this library: No tags from this library for this title. Log in to add tags.
Star ratings
    Average rating: 0.0 (0 votes)
Holdings
Item type Current library Home library Call number Copy number Status Barcode
Thesis قاعة الرسائل الجامعية - الدور الاول المكتبة المركزبة الجديدة - جامعة القاهرة Cai01.12.04.Ph.D.2014.Ad.I (Browse shelf(Opens below)) Not for loan 01010110063829000
CD - Rom مخـــزن الرســائل الجـــامعية - البدروم المكتبة المركزبة الجديدة - جامعة القاهرة Cai01.12.04.Ph.D.2014.Ad.I (Browse shelf(Opens below)) 63829.CD Not for loan 01020110063829000

Thesis (Ph.D.) - Cairo University - Faculty of Science - Department of Biophysics

Purpose: Needs to employ quality assurance (QA) programs that ensure that the system performance meets high expectations consistent with patient care requirements. And to improve target localization accuracy, so that PTV treatment margin can be adjusted or reduced. Materials and Methods: Recently installed an Elekta Synergy linear accelerator with energy 6MV (Elekta, Synergy S, XVI, Crawley, UK) system. Catphan 503 phantom to evaluate CBCT image quality. Ball-Bearing Phantom to evaluate the geometric accuracy of kV and MV isocenters. Quasar{u2122} Penta-Guide Phantom for daily quality assurance (QA). Results: In the present study over a period of 12 months showed that, MV and kV isocenters with mean value of 0.10±0.29mm, -0.07±0.37mm, 0.04±0.28mm and the maximal deviation were 0.45mm, -0.55mm and -0.47mm in x (lateral), y (longitudinal) and z (height) respectively

Issued also as CD

There are no comments on this title.

to post a comment.

Click on an image to view it in the image viewer

Local cover image
Share
Cairo University Libraries Portal Implemented & Customized by: Eng. M. Mohamady Contacts: new-lib@cl.cu.edu.eg | cnul@cl.cu.edu.eg
CUCL logo CNUL logo
© All rights reserved — Cairo University Libraries
CUCL logo
Implemented & Customized by: Eng. M. Mohamady Contact: new-lib@cl.cu.edu.eg © All rights reserved — New Central Library
CNUL logo
Implemented & Customized by: Eng. M. Mohamady Contact: cnul@cl.cu.edu.eg © All rights reserved — Cairo National University Library