A Comparative Study Between The Accuracy Of 10 Mhz And 15 Mhz Ultrasound Probes For Retinal Evaluation In Silicone Oil Filled Globes / by Maha Al Sayed Al Badawi Ibrahim Abdulaaty; Noha Mahmoud Mohamed Khalil, Riham Samy Hanafy Mahmoud Allam, Iman Magdi Eissa

By: Contributor(s): Material type: TextTextLanguage: English Summary language: English, Arabic Producer: 2023Description: 98 pages : illustrations ; 25 cm. + CDContent type:
  • text
Media type:
  • Unmediated
Carrier type:
  • volume
Other title:
  • دراسة مقارنة بين دقة مجسات جهاز الموجات فوق الصوتية ذات تردد ال10 ميجاهيرتز و ال 15 يجاهيرتز في تقييم الشبكية بالعيون الممتلئة بزيت السيليكون [Added title page title]
Subject(s): DDC classification:
  • 617.73
Available additional physical forms:
  • Issued also as CD
Dissertation note: Thesis (Ph.D.)-Cairo University, 2023. Summary: Purpose: To compare 10-MHz and15-MHz B-scan probes regarding detection and localization of retinal detachment (RD) in silicone-oil-filled eyes. Methods: This cross-sectional observational study included 100 eyes (98 patients) -having media opacity precluding fundus examination- scheduled for silicone-oil removal. Patients were examined in the sitting-position using both frequencies one week preoperatively. Longitudinal and transverse scans were taken in primary-gaze, inferior, inferonasal and inferotemporal positions to detect the presence/absence and extent of RD. Patients were sub-grouped according to axial lengths (AXLs), state of silicone emulsification and globe filling. Agreement between sonographic and intraoperative observations was compared. Results: No statistically significant differences were found between 15-MHz and intra-operative findings regarding RD detection (P=0.752) and accurate localization of inferior, inferonasal and inferotemporal RD (P=0.279, 0.606 and 0.599).There were statistically significant differences between 10-MHz and intra-operative findings regarding RD detection and localization (P<0.001).The 15-MHz probe was superior to 10-MHz probe regarding accuracy of RD detection and localization (94% and 47% respectively). The accuracy of 15-MHz probe was 88%, 83% and 85% in detecting and localizing inferior, inferonasal and inferotemporal RD compared to 45%, 60% and 62% with 10-MHz probe. The 15-MHz probe showed better sensitivity while 10-MHz probe showed better accuracy in eyes with short AXLs. The 10-MHz probe showed better sensitivity in patients with sonographic emulsification while15-MHz probe had better sensitivity in detecting vitreoretinal-interface disorders. Conclusion: The 15-MHz B-scan probe is more accurate in detecting and localizing recurrent RD in silicone-oil-filled globes with higher sensitivity in detecting vitreoretinal-interface disorders. Summary: كما أظهرت النتائج أيضاً ز يا دة حساسية مجس الموجات فوق الصوتية ذي ترد د ال 15 ميجاهرتزكلما قل الطول المحوري للعين ؛ حي ث كانت أعلى في المرضى الذين يعانون من طول النظرالمحوري عنالمرضى أصحاب طول العين الطبيعي و المرضى الذين يعانون من قصرالنظرالمحوري. في حي ن أظهر مجس الموجات فوق الصوتية ذو تردد ال 10 ميجاهرتز تحسنًا في الدقة و القيمة التنبؤية الإيجابية في المرضى الذين يعانون من طول النظرالمحوري مقارنةً بـالمرضى أصحاب الطول الطبيعي للعينو الذين يعانون من قصرالنظر المحوري. و قد أظهر مجس الموجات فوق الصوتية ذو تردد ال 10 ميجاهرتز نتائجاً أكثر دقة فيما يتعلقبالحساسية و القيمة التنبؤية السلبية في المرضى الذين يعانون من تحلل زيت السيليكون. كما أظهر كلا المجسين نتائجاً أكثر دقة فيما يتعلق بالحساسية و القيمة التنبؤية الإيجابية في المرضى الذين يعانون من عدم اكتمال امتلاء العين بزيت السيليكون. و توصي هذه الدراسة باستخدام مجس الموجات فوق الصوتية ذي تردد ال 15 ميجاهيرتز في تصويرالعيون الممتلئة بزيت السيليكون والذين يعانون من وجود عتامات تعوق الفحص الإكلينيكي لشبكية العين. وينصح بإجراء الفحص في وضع الجلوس للإستفادة من الدقة العالية لهذا المجس في تصويرالجزء الأسفللشبكية العين, حيث كشفت نتائج الدراسة أن دقة مجس الموجات فوق الصوتية ذي تردد ال 15 ميجاهرتز ف يتشخي ص الانفصال الشبكي كان ت بنسبة 94٪ مقارنة بدقة لا تتجاوز ال 47٪ لمجس الموجات فوق الصوتيةذي تردد ال 10 ميجاهرتز.
Tags from this library: No tags from this library for this title. Log in to add tags.
Star ratings
    Average rating: 0.0 (0 votes)
Holdings
Item type Current library Home library Call number Status Barcode
Thesis Thesis قاعة الرسائل الجامعية - الدور الاول المكتبة المركزبة الجديدة - جامعة القاهرة Cai01.11.24.Ph.D.2023.Ma.C. (Browse shelf(Opens below)) Not for loan 01010110088448000

Thesis (Ph.D.)-Cairo University, 2023.

Referernces: pages 89-98.

Purpose: To compare 10-MHz and15-MHz B-scan probes regarding detection and localization of retinal detachment (RD) in silicone-oil-filled eyes.

Methods: This cross-sectional observational study included 100 eyes (98 patients) -having media opacity precluding fundus examination- scheduled for silicone-oil removal. Patients were examined in the sitting-position using both frequencies one week preoperatively. Longitudinal and transverse scans were taken in primary-gaze, inferior, inferonasal and inferotemporal positions to detect the presence/absence and extent of RD. Patients were sub-grouped according to axial lengths (AXLs), state of silicone emulsification and globe filling. Agreement between sonographic and intraoperative observations was compared.

Results: No statistically significant differences were found between 15-MHz and intra-operative findings regarding RD detection (P=0.752) and accurate localization of inferior, inferonasal and inferotemporal RD (P=0.279, 0.606 and 0.599).There were statistically significant differences between 10-MHz and intra-operative findings regarding RD detection and localization (P<0.001).The 15-MHz probe was superior to 10-MHz probe regarding accuracy of RD detection and localization (94% and 47% respectively). The accuracy of 15-MHz probe was 88%, 83% and 85% in detecting and localizing inferior, inferonasal and inferotemporal RD compared to 45%, 60% and 62% with 10-MHz probe. The 15-MHz probe showed better sensitivity while 10-MHz probe showed better accuracy in eyes with short AXLs. The 10-MHz probe showed better sensitivity in patients with sonographic emulsification while15-MHz probe had better sensitivity in detecting vitreoretinal-interface disorders.
Conclusion: The 15-MHz B-scan probe is more accurate in detecting and localizing recurrent RD in silicone-oil-filled globes with higher sensitivity in detecting vitreoretinal-interface disorders.

كما أظهرت النتائج أيضاً ز يا دة حساسية مجس الموجات فوق الصوتية ذي ترد د ال 15 ميجاهرتزكلما قل الطول المحوري للعين ؛ حي ث كانت أعلى في المرضى الذين يعانون من طول النظرالمحوري عنالمرضى أصحاب طول العين الطبيعي و المرضى الذين يعانون من قصرالنظرالمحوري.

في حي ن أظهر مجس الموجات فوق الصوتية ذو تردد ال 10 ميجاهرتز تحسنًا في الدقة و القيمة التنبؤية الإيجابية في المرضى الذين يعانون من طول النظرالمحوري مقارنةً بـالمرضى أصحاب الطول الطبيعي للعينو الذين يعانون من قصرالنظر المحوري.

و قد أظهر مجس الموجات فوق الصوتية ذو تردد ال 10 ميجاهرتز نتائجاً أكثر دقة فيما يتعلقبالحساسية و القيمة التنبؤية السلبية في المرضى الذين يعانون من تحلل زيت السيليكون.

كما أظهر كلا المجسين نتائجاً أكثر دقة فيما يتعلق بالحساسية و القيمة التنبؤية الإيجابية في المرضى الذين يعانون من عدم اكتمال امتلاء العين بزيت السيليكون.

و توصي هذه الدراسة باستخدام مجس الموجات فوق الصوتية ذي تردد ال 15 ميجاهيرتز في تصويرالعيون الممتلئة بزيت السيليكون والذين يعانون من وجود عتامات تعوق الفحص الإكلينيكي لشبكية العين. وينصح بإجراء الفحص في وضع الجلوس للإستفادة من الدقة العالية لهذا المجس في تصويرالجزء الأسفللشبكية العين, حيث كشفت نتائج الدراسة أن دقة مجس الموجات فوق الصوتية ذي تردد ال 15 ميجاهرتز ف يتشخي ص الانفصال الشبكي كان ت بنسبة 94٪ مقارنة بدقة لا تتجاوز ال 47٪ لمجس الموجات فوق الصوتيةذي تردد ال 10 ميجاهرتز.

Issued also as CD

Text in English and abstract in Arabic & English.