header
Local cover image
Local cover image
Image from OpenLibrary

A Comparison Of Passive Fit Between Conventional And Digital Impression Techniques For An All-On-6 Maxillary Framework : An Invitro Study / By Hasnaa El-Tablawy; Under the supervision of Dr.Amal Kaddah, Dr. Mohamed Farouk, Dr. Nouran Abd Elnaby

By: Contributor(s): Material type: TextTextLanguage: English Summary language: English, Arabic Producer: 2023Description: 121 pages : illustrations ; 25 cm. + CDContent type:
  • text
Media type:
  • Unmediated
Carrier type:
  • volume
Other title:
  • مقارنة بين التوافق السلبي بين تقنيات الانطباع التقليدية والرقمية لإطار شامل للفك العلوي على 6 غرسات : دراسه معمليه [Added title page title]
Subject(s): DDC classification:
  • 617.6
Available additional physical forms:
  • Issued also as CD
Dissertation note: Thesis (M.Sc.) -Cairo University, 2023. Summary: Title of the research in English: A comparison of passive fit between conventional and digital impression techniques for an all-on-6 maxillary framework: An in-vitro study. Aim of performing the research: To compare the passive fit of an implant supported framework fabricated using conventional open tray impression versus the digital impression using intraoral scanner for 6 parallel implants restoring edentulous maxilla. Scientific procedures: Six implants were installed in an epoxy edentulous cast using trial denture base set up of teeth. All implants were installed using dental surveyor to ensure parallelism of all 6 implants, installed in central, canine, and second premolar areas bilaterally. In this invitro study ten frameworks were fabricated. In Group 1: Five Splinted open tray conventional Impressions were carried out for all the six installed implants; each impression was poured in a conventional manner to obtain a master cast. This master cast was used for fabrication of a casted superstructure framework. Group 2: Scan bodies were screwed to all of the installed implants, and five digital impressions using an intra-oral scanner were carried out. The STL files of the five impressions were used to fabricate 5 milled superstructure framework using exocad software. Passive fit of all frameworks fabricate in the two groups was evaluated in the present invitro study using two methods; the first method is the Sheffield test and that would be assessed as passive or non-passive, and the second method was measuring the gap distance using stereomicroscope when all screws were fully tightened and when only the most distal implant was fully tightened in each implant in each constructed framework. Outcome measures: There was no statistically significant difference between the two groups when all screws were fully tightened, while there was statistical significance difference between the two groups when only the most distal implant was fully tightened, results showed more passive fit in casted group compared to milled group. Conclusion: From the results of the current study, we concluded that: Absolute passive fit cannot be achieved regardless of the type of material and technique used. Further evaluation of the reliability and accuracy of the digital impression approach must be undertaken in both laboratory and clinical settings. When evaluating the passive fit of an implant supported full arch maxillary framework fabricated using conventional and digital technique, there was a statistically significant higher vertical gap distance for the digitally fabricated milled framework when compared to the conventional casted frameworkSummary: الهدف من إجراء البحث : مقارنة التوافق السلبي للإطار المدعوم بغرسات الاسنان المصنوع باستخدام انطباع تقليدي مقابل الانطباع الرقمي باستخدام ماسح ضوئي داخل الفم لـ 6 غرسات متوازية لاستعادة الفك العلوي عديم الاسنان. الاجراءات العلميه : تم تثبيت ستة غرسات في قالب إيبوكسي عديم الأسنان باستخدام قاعدة أسنان تجريبية. تم تثبيت جميع الغرسات باستخدام مساح الأسنان لضمان التوازي لجميع الغرسات الستة ، المثبتة في المناطق المركزية ، والكلاب ، والضاحك الثاني بشكل ثنائي في هذه الدراسة تم تصنيع عشرة أطر. في المجموعة 1: تم تنفيذ الانطباعات التقليدية الخمس ذات الدرج المفتوح لجميع الغرسات الست المثبتة ، وتم سكب كل انطباع بطريقة تقليدية للحصول على قالب رئيسي. تم استخدام هذا المصبوب الرئيسي لتصنيع هيكل فوقي مصبوب. المجموعة 2: تم ربط أجسام المسح بجميع الغرسات المثبتة ، وتم إجراء خمس طبعات رقمية باستخدام ماسح ضوئي داخل الفم من نوع ميديت اي 500. تم استخدام ملفات STL للطبعات الخمسة لتصنيع 5 إطار عمل للبنية الفوقية باستخدام برنامج الاكسوكاد. تم تقييم التوافق السلبي لجميع الأطر المصنعة في المجموعتين في الدراسه المعمليه الحالية باستخدام طريقتين ؛ الطريقة الأولى هي اختبار شيفيلد والذي سيتم تقييمه على أنه سلبي أو غير سلبي ، والطريقة الثانية كانت قياس مسافة الفجوة باستخدام المكروسكوب المجسم عندما تكون جميع المسامير مربوطه بالكامل وعندما يتم ربط الزرعة الأبعد فقط في كل غرسة في كل إطار مبني. مقاييس النتائج : لم يكن هناك فرق ذو دلالة إحصائية بين المجموعتين عندما تم ربط جميع المسامير بشكل كامل ، بينما كان هناك فرق دلالة إحصائية بين المجموعتين عندما تم تشديد الزرع الأبعد فقط ، وأظهرت النتائج ملاءمة أكثر سلبية في المجموعة المصبوبة مقارنة بالمجموعة المصنوعه باستخدام الماسح الضوئي الخلاصه : من نتائج الدراسة الحالية يمكننا ان نلخص ما يلي: 1. لا يمكن تحقيق التوافق السلبي المطلق بغض النظر عن التقنية المستخدمة. 2. قد لا يكون نهج الانطباع الرقمي الذي يستخدم الماسحات الضوئية داخل الفم دقيقًا مثل انطباعات مستوى الغرسة التقليدية لإنشاء ملاءمة مقبولة في عمليات تصنيع إطار الزرع طويل المسافه. 3. أظهرت الأطر المصبوبة ملاءمة سلبية أفضل مقارنة بالأطر المصنوعه باستخدام الماسح الضوئي
Tags from this library: No tags from this library for this title. Log in to add tags.
Star ratings
    Average rating: 0.0 (0 votes)
Holdings
Item type Current library Home library Call number Status Barcode
Thesis Thesis قاعة الرسائل الجامعية - الدور الاول المكتبة المركزبة الجديدة - جامعة القاهرة Cai01.09.13.M.Sc.2023.Ha.C. (Browse shelf(Opens below)) Not for loan 01010110088481000
Browsing المكتبة المركزبة الجديدة - جامعة القاهرة shelves Close shelf browser (Hides shelf browser)
No cover image available
No cover image available
No cover image available
No cover image available
No cover image available
No cover image available
No cover image available
Cai01.09.13.M.Sc.2023.Al.B Bone marrow aspirate concentrate (BMAC) on 3D printed Polycaprolactone (PCL) scaffold for ridge augmentation of horizontal alveolar defects in aesthetic zone : Clinical case series study / Cai01.09.13.M.Sc.2023.Al.F. Feasibility Of Patient- Specific Titanium Mesh Via A Transantral Approach For Repair Of Orbital Floor Fractures: A Case Series Study / Cai01.09.13.M.Sc.2023.Al.P. Papilla index & pink esthetic score around customized vs standard healing abutment in immediate implant cases : A randomized controlled trial / Cai01.09.13.M.Sc.2023.Ha.C. A Comparison Of Passive Fit Between Conventional And Digital Impression Techniques For An All-On-6 Maxillary Framework : An Invitro Study / Cai01.09.13.M.Sc.2023.Ka.A Assessment of retention of ot equator smart box attachment system in mandibular implant-retained overdenture with angulated implant versus straight angled implan : An-in vitro study / Cai01.09.13.M.Sc.2023.Ma.A Assessment of vertical ridge Augmentation for posterior mandible using computer guided onlay versus free hand onlay grafting procedures : a randomized clinical split mouth trial / Cai01.09.13.M.Sc.2023.Ma.E. Evaluation Of Surgical Correction Of Anterior Vertical Maxillary Excess (Gummy Smile) By Compression Osteogenesis : Case Series Study /

Thesis (M.Sc.) -Cairo University, 2023.

Bibliography: pages 94-112.

Title of the research in English:
A comparison of passive fit between conventional and digital impression techniques for an all-on-6 maxillary framework: An in-vitro study.
Aim of performing the research:
To compare the passive fit of an implant supported framework fabricated using conventional open tray impression versus the digital impression using intraoral scanner for 6 parallel implants restoring edentulous maxilla.
Scientific procedures:
Six implants were installed in an epoxy edentulous cast using trial denture base set up of teeth. All implants were installed using dental surveyor to ensure parallelism of all 6 implants, installed in central, canine, and second premolar areas bilaterally.
In this invitro study ten frameworks were fabricated.
In Group 1: Five Splinted open tray conventional Impressions were carried out for all the six installed implants; each impression was poured in a conventional manner to obtain a master cast. This master cast was used for fabrication of a casted superstructure framework.
Group 2: Scan bodies were screwed to all of the installed implants, and five digital impressions using an intra-oral scanner were carried out. The STL files of the five impressions were used to fabricate 5 milled superstructure framework using exocad software.
Passive fit of all frameworks fabricate in the two groups was evaluated in the present invitro study using two methods; the first method is the Sheffield test and that would be assessed as passive or non-passive, and the second method was measuring the gap distance using stereomicroscope when all screws were fully tightened and when only the most distal implant was fully tightened in each implant in each constructed framework.
Outcome measures:
There was no statistically significant difference between the two groups when all screws were fully tightened, while there was statistical significance difference between the two groups when only the most distal implant was fully tightened, results showed more passive fit in casted group compared to milled group.
Conclusion:
From the results of the current study, we concluded that:
Absolute passive fit cannot be achieved regardless of the type of material and technique used. Further evaluation of the reliability and accuracy of the digital impression approach must be undertaken in both laboratory and clinical settings.
When evaluating the passive fit of an implant supported full arch maxillary framework fabricated using conventional and digital technique, there was a statistically significant higher vertical gap distance for the digitally fabricated milled framework when compared to the conventional casted framework

الهدف من إجراء البحث :
مقارنة التوافق السلبي للإطار المدعوم بغرسات الاسنان المصنوع باستخدام انطباع تقليدي مقابل الانطباع الرقمي باستخدام ماسح ضوئي داخل الفم لـ 6 غرسات متوازية لاستعادة الفك العلوي عديم الاسنان.
الاجراءات العلميه :
تم تثبيت ستة غرسات في قالب إيبوكسي عديم الأسنان باستخدام قاعدة أسنان تجريبية. تم تثبيت جميع الغرسات باستخدام مساح الأسنان لضمان التوازي لجميع الغرسات الستة ، المثبتة في المناطق المركزية ، والكلاب ، والضاحك الثاني بشكل ثنائي
في هذه الدراسة تم تصنيع عشرة أطر.
في المجموعة 1: تم تنفيذ الانطباعات التقليدية الخمس ذات الدرج المفتوح لجميع الغرسات الست المثبتة ، وتم سكب كل انطباع بطريقة تقليدية للحصول على قالب رئيسي. تم استخدام هذا المصبوب الرئيسي لتصنيع هيكل فوقي مصبوب.
المجموعة 2: تم ربط أجسام المسح بجميع الغرسات المثبتة ، وتم إجراء خمس طبعات رقمية باستخدام ماسح ضوئي داخل الفم من نوع ميديت اي 500. تم استخدام ملفات STL للطبعات الخمسة لتصنيع 5 إطار عمل للبنية الفوقية باستخدام برنامج الاكسوكاد.
تم تقييم التوافق السلبي لجميع الأطر المصنعة في المجموعتين في الدراسه المعمليه الحالية باستخدام طريقتين ؛ الطريقة الأولى هي اختبار شيفيلد والذي سيتم تقييمه على أنه سلبي أو غير سلبي ، والطريقة الثانية كانت قياس مسافة الفجوة باستخدام المكروسكوب المجسم عندما تكون جميع المسامير مربوطه بالكامل وعندما يتم ربط الزرعة الأبعد فقط في كل غرسة في كل إطار مبني.
مقاييس النتائج :
لم يكن هناك فرق ذو دلالة إحصائية بين المجموعتين عندما تم ربط جميع المسامير بشكل كامل ، بينما كان هناك فرق دلالة إحصائية بين المجموعتين عندما تم تشديد الزرع الأبعد فقط ، وأظهرت النتائج ملاءمة أكثر سلبية في المجموعة المصبوبة مقارنة بالمجموعة المصنوعه باستخدام الماسح الضوئي
الخلاصه :
من نتائج الدراسة الحالية يمكننا ان نلخص ما يلي:
1. لا يمكن تحقيق التوافق السلبي المطلق بغض النظر عن التقنية المستخدمة.
2. قد لا يكون نهج الانطباع الرقمي الذي يستخدم الماسحات الضوئية داخل الفم دقيقًا مثل انطباعات مستوى الغرسة التقليدية لإنشاء ملاءمة مقبولة في عمليات تصنيع إطار الزرع طويل المسافه.
3. أظهرت الأطر المصبوبة ملاءمة سلبية أفضل مقارنة بالأطر المصنوعه باستخدام الماسح الضوئي

Issued also as CD