Accuracy of linear measurements of the planmeca promax 3D MID® cone beam computed tomography machine in different head orientations : A diagnostic accuracy study / Nada Saeed Abdelhameed Mustafa ; Supervised Sahar Hosny Eldessouky , Dina Mohamed Elbeshlawy
Material type:
- فى اوضاع مختلفه للرأس PlanmecaPromax 3D MID® دقة القياسات الخطيه للتصويرالمقطعي بالحاسوب مخروطي الشعاع لجهاز : دراسة دقة تشخيصية [Added title page title]
- Issued also as CD
Item type | Current library | Home library | Call number | Copy number | Status | Barcode | |
---|---|---|---|---|---|---|---|
![]() |
قاعة الرسائل الجامعية - الدور الاول | المكتبة المركزبة الجديدة - جامعة القاهرة | Cai01.09.05.M.Sc.2017.Na.A (Browse shelf(Opens below)) | Not for loan | 01010110075544000 | ||
![]() |
مخـــزن الرســائل الجـــامعية - البدروم | المكتبة المركزبة الجديدة - جامعة القاهرة | Cai01.09.05.M.Sc.2017.Na.A (Browse shelf(Opens below)) | 75544.CD | Not for loan | 01020110075544000 |
Browsing المكتبة المركزبة الجديدة - جامعة القاهرة shelves Close shelf browser (Hides shelf browser)
Thesis (M.Sc.) - Cairo University - Faculty of Oral and Dental Medicine - Department of Oral (Radiology and Diagnosis)
Nowadays, Cone beam computed tomography is widely used for both diagnosis and treatment planning in many dental fields providing precise three dimensional information about dental and maxillofacial structures with almost infinite number of image display options. Obtaining linear measurements is one of the very helpful facilities provided by all CBCT software and the dentists rely on such measurements for taking the right decision especially as regard implant placement, endodontic treatment as well as in cases requiring orthodontic intervention. The lack of accuracy of these linear measurements may lead to serious problems which might endanger the patient or adversely affect the treatment outcome. The patient{u2019}s head position during the CBCT scanning could deviate from the ideal one predetermined by each CBCT machine manufacturers. It is therefore important not only to assess the accuracy of linear measurements obtained from CBCT scans in standard scanning position, but also to examine the effect of improper head positioning on the accuracy of these measurements. Some controversies concerning the influence of changing the head positions on the accuracy of CBCT linear measurements were found on the previous studies. This study aimed to assess the accuracy of linear measurements of PlanmecaPromax 3D MID® CBCT machine in both standard and different head positions. Eleven dry human skulls were included in the study. Guttapercha markers were placed to define the distances to be measured in occluso-apical, bucco-lingual and mesio-distal dimensions. These 3 dimensions were measured in anterior, bilateral premolars and bilateral molars regions for each jaw and the skulls were imaged in standard and four different head positions (forward tilt, backward tilt, lateral tilt and rotation). Real measurements were taken using a digital caliper and CBCT measurements were performed by two radiologists separately