Image from OpenLibrary

A vendor-neutral strain elastography technique: experimental validation / by Mohamed Salah El Din Ahmed ; Supervisors Prof. Dr. Ahmed M Ehab Mahmoud.

By: Contributor(s): Material type: TextLanguage: English Summary language: English, Arabic Producer: 2025Description: 300 pages : illustrations ; 30 cm. + CDContent type:
  • text
Media type:
  • Unmediated
Carrier type:
  • volume
Other title:
  • تقنية تصوير المرونة غير المتحيز : التحقق التجريبي [Added title page title]
Subject(s): DDC classification:
  • 610.28
Available additional physical forms:
  • Issues also as CD.
Dissertation note: Thesis (M.Sc)-Cairo University, 2025. Summary: Ultrasound (US) elastography is a crucial non-invasive technique for assessing tissue stiffness, aiding in diagnosing conditions like liver disease and tumors. This study validates US-Elasto, a manufacturer-independent add-on module designed to integrate elastography into conventional B-mode ultrasound systems. Tested on 10 US machines from 7 manufacturers using a breast elastography phantom, US-Elasto employed a novel displacement tracking technique to generate strain images, revealing a strain ratio of 2.52±0.29 between stiffer masses and background tissue. Clinical validation with 81 breast scans from 56 patients demonstrated strong agreement (r = 0.937, R² = 0.88) with commercial elastography systems. These results confirm US-Elasto’s accuracy and potential as a cost-effective, vendor-neutral solution, expanding access to advanced elastography for broader clinical use.Summary: تعد المرونة بالموجات فوق الصوتية تقنية غير جراحية مهمة لتقييم صلابة الأنسجة، مما يساعد في تشخيص حالات مثل أمراض الكبد والأورام. تهدف هذه الدراسة إلى التحقق من صحة تقنية تصوير المرونة غير المتحيز, وهو وحدة إضافية مستقلة عن الشركة المصنعة تتيح دمج المرونة مع أنظمة الموجات فوق الصوتية التقليدية. تم اختبارها على 10 أجهزة من 7 شركات مصنِّعة باستخدام نموذج مرونة للثدي، حيث تم تطبيق تقنية تتبع الإزاحة لحساب الإجهاد، مما أظهر نسبة إجهاد 2.52±0.29 بين الكتل الأكثر صلابة والأنسجة المحيطة. شملت التجارب السريرية 81 فحصًا للثدي لـ 56 مريضة، وأظهرت توافقًا قويًا مع أنظمة المرونة التجارية. تؤكد هذه النتائج دقة وإمكاناته الوحده كحل منخفض التكلفة ومستقل عن الشركات المصنعة، مما يوسع نطاق الوصول إلى المرونة في التطبيقات الطبية.
Tags from this library: No tags from this library for this title. Log in to add tags.
Star ratings
    Average rating: 0.0 (0 votes)
Holdings
Item type Current library Home library Call number Status Barcode
Thesis قاعة الرسائل الجامعية - الدور الاول المكتبة المركزبة الجديدة - جامعة القاهرة Cai01.13.03.M.Sc.2025.Mo.V (Browse shelf(Opens below)) Not for loan 01010110092781000

Thesis (M.Sc)-Cairo University, 2025.

Bibliography: pages 68-73.

Ultrasound (US) elastography is a crucial non-invasive technique for assessing tissue stiffness, aiding in diagnosing conditions like liver disease and tumors. This study validates US-Elasto, a manufacturer-independent add-on module designed to integrate elastography into conventional B-mode ultrasound systems. Tested on 10 US machines from 7 manufacturers using a breast elastography phantom, US-Elasto employed a novel displacement tracking technique to generate strain images, revealing a strain ratio of 2.52±0.29 between stiffer masses and background tissue. Clinical validation with 81 breast scans from 56 patients demonstrated strong agreement (r = 0.937, R² = 0.88) with commercial elastography systems. These results confirm US-Elasto’s accuracy and potential as a cost-effective, vendor-neutral solution, expanding access to advanced elastography for broader clinical use.

تعد المرونة بالموجات فوق الصوتية تقنية غير جراحية مهمة لتقييم صلابة الأنسجة، مما يساعد في تشخيص حالات مثل أمراض الكبد والأورام. تهدف هذه الدراسة إلى التحقق من صحة تقنية تصوير المرونة غير المتحيز, وهو وحدة إضافية مستقلة عن الشركة المصنعة تتيح دمج المرونة مع أنظمة الموجات فوق الصوتية التقليدية. تم اختبارها على 10 أجهزة من 7 شركات مصنِّعة باستخدام نموذج مرونة للثدي، حيث تم تطبيق تقنية تتبع الإزاحة لحساب الإجهاد، مما أظهر نسبة إجهاد 2.52±0.29 بين الكتل الأكثر صلابة والأنسجة المحيطة. شملت التجارب السريرية 81 فحصًا للثدي لـ 56 مريضة، وأظهرت توافقًا قويًا مع أنظمة المرونة التجارية. تؤكد هذه النتائج دقة وإمكاناته الوحده كحل منخفض التكلفة ومستقل عن الشركات المصنعة، مما يوسع نطاق الوصول إلى المرونة في التطبيقات الطبية.

Issues also as CD.

Text in English and abstract in Arabic & English.

There are no comments on this title.

to post a comment.
Share
Cairo University Libraries Portal Implemented & Customized by: Eng. M. Mohamady Contacts: new-lib@cl.cu.edu.eg | cnul@cl.cu.edu.eg
CUCL logo CNUL logo
© All rights reserved — Cairo University Libraries
CUCL logo
Implemented & Customized by: Eng. M. Mohamady Contact: new-lib@cl.cu.edu.eg © All rights reserved — New Central Library
CNUL logo
Implemented & Customized by: Eng. M. Mohamady Contact: cnul@cl.cu.edu.eg © All rights reserved — Cairo National University Library