Retention of mandibular implant supported overdentures using CM-LOC and locator attachment systems : A randomized in vitro study / by Omar Hashem Kamal Mohamed ; Supervisor Dr. Nouran Abd Al Naby, Dr. Nada Sherin El Khourazaty.

By: Contributor(s): Material type: TextLanguage: English Summary language: English, Arabic Producer: 2020Description: 101 pages : illustrations ; 25 cm. + CDContent type:
  • text
Media type:
  • Unmediated
Carrier type:
  • volume
Other title:
  • استبقاء الاطقم الكاملة المدعومة بغرسات فى الفك السفلى باستخدام Locator و CM-LOC من الوصلات المستخدمة لاستبقاء دراسة معملية [Added title page title]
Subject(s): DDC classification:
  • 617.69
Available additional physical forms:
  • Issues also as CD.
Dissertation note: Thesis (M.Sc.)-Cairo University, 2020. Summary: Aim: Implant supported overdentures have proved to be a better treatment option compared to conventional dentures in treating edentulous jaws. The present study was carried out to compare the retention of two different types of attachments used to retain single implant supported mandibular overdenture using universal testing machine after mechanical aging that was done by chewing simulator. The two attachments used in this study were Locator and Cendres and Meteaux Locator (CM-LOC). Methodology: Ten epoxy resin models of completely edentulous mandible were constructed and prepared to receive a single implant in the midline area. Ten overdentures were constructed with steel rods fixed to the denture using auto polymerized acrylic resin at the geometric center, two attachment systems; Locator attachmrent with nylon matrix and Cendres &Meteaux Locator (CM-LOC) with a PEKK matrix were investigated, with each type of attachment, the pickup procedure consisted of attaching matrix to the fitting surface of the denture with self-cure acrylic resin. All of the overdentures were subjected to different chewing cycles 500 cycles = (3 month), 1000 cycles = (6 month), 1500 cycles = (9 month) and 2000 cycles = (12 month) , then retention was evaluated using universal testing machine.. Results: The Locator attachment have shown a decrease in retention when subjected to different chewing cycles (p=0.001), with pairwise comparison showing a statistically significant difference between baseline and 2000 cycles ( p=0.011), between 500 cycle and 2000 cycle ( p=0.024). While for the CM-LOC group there was a decrease in retention when subjected to 500, 1000, 1500 cylces ( p=0.001), and with a slight increase in retention after 2000 cycles. With pairwise comparisons there was statistically significant difference at 500 cycles = 3 month and 1500 cycles (9 month). Conclusion: Within in limitation of the present study in vitro study it can be concluded that both the Locator and CM-LOC attachment have shown acceptable retention values to support single implant mandibular retained overdentures. However both attachments have shown a decrease in retention when subjected to different chewing cycles , The CM-LOC attachment has shown significant better retention when compared to Locator attachment after 500 cycles . Summary: اجريت هذه الرسالة لمقارنة القدرة الاحتياطية لنوعين مختلفين من الملحقات المستخدمة للاحتفاظ في الاطقم المدعومة بالزرعة الواحدة بواسطة ألة اختبار علمية بعد التقدم الميكانيكي الذي تم بواسطة جهاز محاكاة المضغ. بينما كانت مجموعة التدخل هي مجموعة CM-LOC وكانت مجموعة التحكم هيLOCATOR تم بناء نموذج ايبوكسي للفك السفلي عديم الاسنان تماماً وتجهيزه لتلقي زرعة واحدة في منطقة خط الوسط. تم بناء عشرة أطقم مفردة خمسة تحتوي على مصفوفة من النايلون وخمسة أخرى تحتوي على مصفوفة من PEEK ثم يتم قياس قيم الاحتفاظ عند خط الاساس باستخدام ألة اختبارعالمية ثم تعرضت الأطقم في كل مجموعة لدورات مضغ مختلفة 500 و 1000 و 1500 و 2000 دورة. تم قياس الاستبقاء طوال هذه الدورات لكلا المجموعتين.
Tags from this library: No tags from this library for this title. Log in to add tags.
Star ratings
    Average rating: 0.0 (0 votes)
Holdings
Item type Current library Home library Call number Status Barcode
Thesis قاعة الرسائل الجامعية - الدور الاول المكتبة المركزبة الجديدة - جامعة القاهرة Cai01.09.09.M.Sc.2020.Om.R (Browse shelf(Opens below)) Not for loan 01010110093150000

Thesis (M.Sc.)-Cairo University, 2020.

Bibliography: pages 86-100.

Aim: Implant supported overdentures have proved to be a better treatment option compared to
conventional dentures in treating edentulous jaws. The present study was carried out to compare the
retention of two different types of attachments used to retain single implant supported mandibular
overdenture using universal testing machine after mechanical aging that was done by chewing simulator.
The two attachments used in this study were Locator and Cendres and Meteaux Locator (CM-LOC).



Methodology: Ten epoxy resin models of completely edentulous mandible were constructed and
prepared to receive a single implant in the midline area. Ten overdentures were constructed with steel
rods fixed to the denture using auto polymerized acrylic resin at the geometric center, two attachment
systems; Locator attachmrent with nylon matrix and Cendres &Meteaux Locator (CM-LOC) with a
PEKK matrix were investigated, with each type of attachment, the pickup procedure consisted of
attaching matrix to the fitting surface of the denture with self-cure acrylic resin. All of the overdentures
were subjected to different chewing cycles 500 cycles = (3 month), 1000 cycles = (6 month), 1500
cycles = (9 month) and 2000 cycles = (12 month) , then retention was evaluated using universal testing
machine..
Results: The Locator attachment have shown a decrease in retention when subjected to different
chewing cycles (p=0.001), with pairwise comparison showing a statistically significant difference
between baseline and 2000 cycles ( p=0.011), between 500 cycle and 2000 cycle ( p=0.024).
While for the CM-LOC group there was a decrease in retention when subjected to 500, 1000, 1500
cylces ( p=0.001), and with a slight increase in retention after 2000 cycles. With pairwise comparisons
there was statistically significant difference at 500 cycles = 3 month and 1500 cycles (9 month).
Conclusion: Within in limitation of the present study in vitro study it can be concluded that both the
Locator and CM-LOC attachment have shown acceptable retention values to support single implant
mandibular retained overdentures. However both attachments have shown a decrease in retention when
subjected to different chewing cycles , The CM-LOC attachment has shown significant better retention
when compared to Locator attachment after 500 cycles .

اجريت هذه الرسالة لمقارنة القدرة الاحتياطية لنوعين مختلفين من الملحقات المستخدمة للاحتفاظ في الاطقم المدعومة بالزرعة الواحدة بواسطة ألة اختبار علمية بعد التقدم الميكانيكي الذي تم بواسطة جهاز محاكاة المضغ.
بينما كانت مجموعة التدخل هي مجموعة CM-LOC وكانت مجموعة التحكم هيLOCATOR
تم بناء نموذج ايبوكسي للفك السفلي عديم الاسنان تماماً وتجهيزه لتلقي زرعة واحدة في منطقة خط الوسط. تم بناء عشرة أطقم مفردة خمسة تحتوي على مصفوفة من النايلون وخمسة أخرى تحتوي على مصفوفة من PEEK
ثم يتم قياس قيم الاحتفاظ عند خط الاساس باستخدام ألة اختبارعالمية ثم تعرضت الأطقم في كل مجموعة لدورات مضغ مختلفة 500 و 1000 و 1500 و 2000 دورة. تم قياس الاستبقاء طوال هذه الدورات لكلا المجموعتين.

Issues also as CD.

Text in English and abstract in Arabic & English.

There are no comments on this title.

to post a comment.
Share
Cairo University Libraries Portal Implemented & Customized by: Eng. M. Mohamady Contacts: new-lib@cl.cu.edu.eg | cnul@cl.cu.edu.eg
CUCL logo CNUL logo
© All rights reserved — Cairo University Libraries
CUCL logo
Implemented & Customized by: Eng. M. Mohamady Contact: new-lib@cl.cu.edu.eg © All rights reserved — New Central Library
CNUL logo
Implemented & Customized by: Eng. M. Mohamady Contact: cnul@cl.cu.edu.eg © All rights reserved — Cairo National University Library